德尔格Evita4呼吸机故障代码表(维修技术支持配件供应 13414141187姜)
2.1Errorcode02SafetySoftwareThe service codes 02.00.. to 02.22.. are recognised on theCPU 68332 PCB; theyfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB3F.tmp.pngcorrespond to02.40.. to 02.62.. as recognised on thePneumatic Controller PCB.
FaultItemCause/Remedy
02.00.001si0_Err, New start, cause incorrect, resetMonitoring of cold/warm start storage in RAM onCPU 68332 PCB. Cause: Software error onCPU 68332 PCB
02.00.002si0_Err, Fast data chksum error, resetData packets betweenCPU 68332PCB and Pneumatic Controller PCBcannot be transferred in full.CPU68332 PCB or Pneumatic ControllerPCB defective.
02.00.003si0_Err, Safety software of partner CPU not started, resetAfter starting, it was not possible within10 seconds to establish synchronisation betweenCPU 68332 PCB and Pneumatic Controller PCB. Cause:CPU 68332 PCB or PneumaticController PCB defective.
02.00.004si0_Err, Input data, chksum error, resetData packets betweenCPU 68332PCB and Pneumatic Controller PCBcannot be transferred in full.CPU68332 PCB or Pneumatic ControllerPCB defective.
02.00.005si0_Err, Vector for NMIincorrect, resetRAM or SW error onCPU 68332 PCB
02.00.020si0_Err, Si boot, case incorrect, resetRAM or SW error onCPU 68332 PCBor Pneumatic Controller PCB.
5664.500/800 Evita 4/Evita 2 dura 01.2002Repair Instructions – 4
file:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB4F.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB50.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB51.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB52.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB53.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB54.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB55.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB56.tmp.jpg
FaultItemCause/Remedy
02.01.001si1_ Err, Adjuster has incorrect pixel sum, X alarm, reset:Screen content (with Evita 2 dura also 7- segment displays) is read out and compared to request. Cause: Hardware or SW error on Graphics ControllerPCB (Evita 4)/Front Panel PCB(Evita 2 dura) orCPU 68332 PCB or Pneumatic Controller PCB. If display plausible, then SW error; otherwise high probability of hardware error on Graphics Controller PCB (Evita 4) or Front Panel PCB (Evita 2 dura). Table for X:
1 Paw high adjuster selected2 Paw high arrow selected3 MV min adjuster selected4 MV min arrow selected5 O2 adjuster selected6 Last adjuster plus one7 Paw high adjuster not selected8 Paw high arrow not selected9 MV min adjuster not selected10 MV min arrow not selected11 O2 adjuster not selected12 Paw high line 30 top13 Pressure scale values top 3014 Paw high line 60 top15 Pressure scale values top 6016 Paw high line 100 top17 Pressure scale values top 10018 Paw high line 30 bottom19 Pressure scale values bottom 3020 Paw high line 60 bottom21 Pressure scale values bottom 6022 Paw high line 100 bottom23 Pressure scale values bottom 10024 Paw high line 30 bargraph25 Pressure scale values bargraph 3026 Paw high line 60 bargraph27 Pressure scale values bargraph 6028 Paw high line 100 bargraph29 Pressure scale values bargraph10030 Paw high display 131 Paw high display 232 MV min display 133 MV min display 234 Cold start screen
file:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB67.tmp.png5664.500/800 Evita 4/Evita 2 dura 01.2002Repair Instructions – 5
file:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB68.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB69.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB6A.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB6B.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB7B.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB7C.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB7D.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB7E.tmp.jpg
FaultItemCause/Remedy
02.01.002si1_Err, Paw high, Si copy, consistency X alarm, resetAlarm values relevant to safety are checked for plausibility.Cause: RAM or SW error onCPU68332 PCB orPneumatic ControllerPCB.
02.01.003si1_Err, MV min, Si copy, consistency X alarm, resetAlarm values relevant to safety are checked for plausibility.Cause: RAM or SW error onCPU68332 PCB or Pneumatic ControllerPCB.
02.01.004si1_Err, O2 adjust, Si copy, consistency X alarm, resetAlarm values relevant to safety are checked for plausibility.Cause: RAM or SW error onCPU68332 PCB or Pneumatic ControllerPCB.
02.01.005si1_Err, Paw high, invalid value, resetAlarm values relevant to safety are checked for plausibility.Cause: RAM or SW error onCPU68332 PCB or Pneumatic ControllerPCB.
02.01.006si1_Err, MV min, invalid value, resetAlarm values relevant to safety are checked for plausibility.Cause: RAM or SW error onCPU68332 PCB or Pneumatic ControllerPCB.
02.01.007si1_Err, O2 adjustment, invalid value, resetAlarm values relevant to safety are checked for plausibility.Cause: RAM or SW error onCPU68332 PCB or Pneumatic ControllerPCB.
02.01.020si1_Err, Si boot, case incorrect, resetRAM or SW error onCPU 68332 PCBorPneumatic Controller PCB.
02.01.021si1_Err, Xxx is set, case incorrect, resetOnly applies up to SW 1.04. RAM or SW error onCPU 68332 PCB or Pneumatic Controller PCB
02.01.022si1_Err, Control console, status, case incorrect, resetRAM or SW error onCPU 68332 PCBorPneumatic Controller PCB.
file:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB7F.tmp.png5664.500/800 Evita 4/Evita 2 dura 01.2002Repair Instructions – 6
file:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB80.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB91.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB92.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB93.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB94.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB95.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB96.tmp.pngfile:///C:\Users\mqs\AppData\Local\Temp\ksohtml\wpsEB97.tmp.jpg
FaultItemCause/Remedy
02.02.001si2_Err, Identifier has incorrect pixel sum, X alarm, reset专业德尔格呼吸麻醉机维修13414141187姜Screen content (with Evita 2 dura also 7- segment displays) is read out and compared to request. Cause: Hardware or SW error onGraphics ControllerPCB (Evita 4) /Front Panel PCB (Evita2 dura) orCPU 68332 PCB or Pneumatic Controller PCB. If display plausible, then SW error; otherwise high probability of hardware error on Graphics Controller PCB (Evita 4) orFront Panel PCB (Evita 2 dura). Refer to error 02.01.001 for X table.
02.02.002si2_Err, Identifier not announced, X alarm, resetRefer to error 02.02.001.
02.02.003si2_Err, Value has incorrect pixel sum, X alarm, resetRefer to error 02.02.001.
02.02.020si2_Err, Si boot, case incorrect, resetRAM or SW error onCPU 68332 PCBorPneumatic Controller PCB.
02.03.001si3_Err, Keypad input sequence, counter incorrect, resetSW or transmission error fromGraphics Controller PCB (Evita 4) orFront Panel PCB (Evita 2 dura) viaCPU68332 PCB to Pneumatic Controller PCB. Cause: Probably error on Graphics Controller PCB (Evita 4) or Front Panel PCB (Evita 2 dura).
02.03.002si3_Err, Rotary knob button pressed, incorrectly displayed, resetSW or hardware error when reading out keys onGraphics Controller PCB(Evita 4) orFront Panel PCB(Evita 2 dura).
02.03.003si3_Err, Silence button pressed, incorrectly displayed, resetRefer to error 02.03.002.
02.03.004si3_Err, Reset button pressed, incorrectly displayed, resetRefer to error 02.03.002.
02.03.005si3_Err, Extraction button pressed, incorrectly displayed, resetRefer to error 02.03.002.
02.03.006si3_Err, Inspiration hold button pressed, incorrectly displayed, resetRefer to error 02.03.002.
02.03.007si3_Err, Expiration hold button pressed, incorrectly displayed, resetRefer to error 02.03.002.