贵阳罗工 发表于 2014-6-21 21:19:49

费森尤斯血液透析机4008系列维修手册

(维修手册,水路图详见附件)
“V84 faultiness”报警与V84阀的清洗 “V84 faultiness”报警通常是因为V84阀两侧的电极探测到了消毒液的残余,可以通过以下步骤清洗电极和V84阀,以消除故障。 1.将消毒液吸管拔出,放入一个盛有约5升的容器内。2.维修开关上拨,开机进入诊断程序-DIAGNOSTICS3.按以下菜单顺序依次按确认键:DIAGNOSTICSà WRITEOUTPUTS à WRITEDIGIT. OUTPUTS à CPU1:WRDIGIT. OUTP ,按确认键,此时显示:O:CPU1_V244.按file:///C:/Users/ADMINI~1/AppData/Local/Temp/msohtmlclip1/01/clip_image002.gif键找到:O:CPU1_FL_SWITCH_EN ,此时按面板上UF Rate 旁file:///C:/Users/ADMINI~1/AppData/Local/Temp/msohtmlclip1/01/clip_image002.gif键使其显示:1111,使V41阀和浮子开关正常工作。5.按file:///C:/Users/ADMINI~1/AppData/Local/Temp/msohtmlclip1/01/clip_image002.gif键找到:O:CPU1_V26,此时按面板上UF Rate 旁file:///C:/Users/ADMINI~1/AppData/Local/Temp/msohtmlclip1/01/clip_image002.gif键使其显示:1111,使V26阀打开。6. 以同样的方法将 V31,V33,V35,V37,V84 阀打开。7. 按file:///C:/Users/ADMINI~1/AppData/Local/Temp/msohtmlclip1/01/clip_image002.gif键找到:O:CPU1_STOP_EP ,此时按面板上UF Rate 旁file:///C:/Users/ADMINI~1/AppData/Local/Temp/msohtmlclip1/01/clip_image002.gif键使其显示:0000,负压泵开始工作并冲洗V84阀。8.5分钟后关机,将维修开关拨下。 “Upper flow alarm ” 报警与流量的校正 流量报警有时可以通过以下步骤校正流量,以消除故障。 1.关机,维修开关上拨,开机进入校正状态: CALIBRATION。2.将A,B液吸管插入透析液桶内,打开流量 FLOW,等待两分钟。3.按以下菜单顺序依次按确认键:CALIBRATION à CAL. FLOW 300 ml/minà flow(300)=xxx4.逆时针调整电路板LP634的P1电位器,直至显示:flow(300)=1475.顺时针缓慢地调整P1,每调节1/8圈等待15秒,直至显示 flow(300)=xxx, 其值不再是147,而且大于200,表明已调到临界点,在此位置上顺时针调2圈整,流量脉冲调节完毕。6.按file:///C:/Users/ADMINI~1/AppData/Local/Temp/msohtmlclip1/01/clip_image002.gif键直至显示:flow(300)=300左右,按Override键(4008B)或Mute键(4008S)存储。7.依照步骤6同样校正流量500和800。8.关机,将维修开关拨下。